Accela* High Speed LC

  Accela* High Speed LC



Thermo Scientific* Accela High Speed LC provides fast, efficient chromatographic separations over a range of flowrates and pressures for HPLC and UHPLC in a single system using the industry's only quaternary high pressure pump.

   Product Detail


Thermo Scientific chromatography instruments allow users to achieve shorter run times with excellent peak symmetry and superior data quality

Accela is designed to optimize performance of sub-two micron particle columns, providing seamless operation from conventional LC pressures up to 1250 bar for U-HPLC separations of complex matrices.
  • Rapid gradient formation with extremely low delay volumes
  • Maximum application flexibility with HPLC/U-HPLC in a single system
  • Precise flow and gradients with proprietary Force Feedback Control (FFC) technology
  • Streamlined integration across a full range of LC & MS detectors

Accela 1250 Pump
  • Accelerated method development time with quaternary solvent delivery
  • HPLC & U-HPLC applications up to 1250 bar
  • Extremely fast cycle with a 70µL delay volume
  • FFC technology ensures consistent baseline reproducibility
Accela 600 Pump
  • Accelerated method development time with quaternary solvent delivery
  • Greater application flexibility with flow rates up to 5mL/min at 600 bar
  • Minimum gradient equilibration time with a 90µL delay volume
  • FFC technology eliminates the need for pulse dampeners and reduces maintenance costs
Accela Open Autosampler
  • Maximum flexibility with unmatched reliability and robustness
  • Customizable temperature controlled sample formats
  • Integrated Dynamic Load and Wash capability resulting in extremely low carryover (0.003% chlorhexidine)
  • Saves precious bench space with a customizable open frame architecture in a stackable design
Accela Autosampler
  • Optimized for high-throughput applications and sub two micron column particles
  • Maximum reproducibility with isothermal injection and separation
  • Greater application flexibility with multiple sample formats
  • Superior robustness with a diamond coated high pressure valve Accela Photo Diode Array Detector
  • Highest detection sensitivity with proprietary LightPipe technology
  • Designed for U-HPLC separations with a 1cm path length, 2µL LightPipe flowcell
  • Maximum peak resolution with fiberoptic beam shaper
  • Easy maintenance with integrated wavelength validation
Accela Photo Diode Array Detector
  • Highest detection sensitivity with proprietary LightPipe technology
  • Designed for U-HPLC separations with a 1cm path length, 2µL LightPipe flowcell
  • Maximum peak resolution with fiberoptic beam shaper
  • Easy maintenance with integrated wavelength validation
Accela UV/Vis Detector
  • Unlimited applications flexibility with externally mounted flow cells
  • Simultaneous detection at two selectable wavelengths in the UV or the visible range of the spectrum
  • Instrument versatility with two analog outputs
  • Simple maintenance with pre-aligned deuterium and tungsten lamps
Accela Refractive Index (RI) Detector
  • High sensitivity HPLC detection
  • Faster throughput with a low delay volume flowcell
  • Automatic lamp brightness compensation for flowcell stains
  • Optimum sensitivity and selectivity with Thermo Scientific columns

Ideal for:
Forensics, Pharmaceutical, Biodiesel Analysis, Food Safety, Environmental, Forensics, Evidence Analysis, Drug Screening, Metabolite Analysis
 Accela* High Speed LC
 Product# Price  Description  
       
 6005760030      -    Performs HPLC and UHPLC, upper pressure limit 15,000 psi  Select 
 
Related Products
 Product# Product Name Image  
 
IQLAAAGAAU
FAAAMAYB
 
Accela 1250 UHPLC Pump
  Select   
 
IQLAAAGAAU
FAAAMAYC
 
Accela 600 Pump
  Select   
 
IQLAAEGAAP
FADAMAII
 
LTQ FT Ultra Hybrid Mass Spectrometer
  Select   
 
IQLAAEGAAP
FADBMAIJ
 
LTQ Orbitrap Hybrid Mass Spectrometers
  Select   
 
IQLAAEGAAV
FACZMAIK
 
LTQ XL Linear Ion Trap Mass Spectrometer
  Select   
 
IQLAAEGAAV
FADCMAIL
 
LXQ Linear Ion Trap Mass Spectrometer
  Select   
 
IQLAAAGAAW
FAFIMALO
 
MSQ Plus Mass Detector
  Select   
 
IQLAAEGAAV
FACZMAWQ
 
Velos Pro Dual-Pressure Linear Ion Trap
  Select   
 
IQLAAEGABO
FAGVMZZZ/I
NQSOF012
 
Xcalibur
  Select   
Purchase Details
 
Request Quote
Product Contact
 Sales Contact
  Service Contact
 
Literature Request
Postal Mail:
  • Successful HPLC Operation: A Troubleshooting Guide (CD)
  • FREE 2008/2009 Chromatography Columns & Consumables Catalog
  • Electronic Mail:
  • LCQ Fleet Brochure (PDF)
  • U-HPLC in the Analysis of Phenolic Pollutants in Water (PDF)
  • Method Transfer from HPLC to UHPLC for Increased Speed and Resolution (PDF)
  • Accela Autosampler Product Specification (PDF)
  • Accela PDA Detector Product Specification (PDF)
  • Accela High Speed Pump Product Specification (PDF)
  • Accela High Speed LC System Brochure (PDF)
  • Thermo Scientific Sistemas LC Accela
  • Specifications:
    Absorbance Linearity :<5% up to 2.0 AU at 254nm
    Absorbance Non-Linearity :<5% at 2.0 AU at 257nm
    Absorbance Range :0.0005 AUFS to + 3.0 AUFS
    Ambient Environment :4 to 40°C, 5 to 80% relative humidity, (non condensing)
    Analog Outputs :20-bit digital/analog conversion, unattenuated at 10mV/AU,100mV/AU, or 1.0V/AU
    Auto Zero Range :±10 x 10-5 RIU
    Auto Zero Resolution :±10 x 10-9 RIU
    Cell Dimensions :10mm, 15 µL standard analytical
    50mm, 10 µL LightPipe
    10mm, 2 µL LightPipe
    Cell Pressure Range :0 to 1000psi
    Column Oven :5 to 95°C ± 0.5°C selectable in 1°C increments
    Communications :Remote Inputs
    Pump Ready and Inject Hold
    Compositional Accuracy :±0.50%
    Compositional Precision :<0.2% RSD
    Degasser Hold Volume :<500µL per channel
    Degasser Type :Integral, self-adjusting, 4 channel vacuum membrane (PTFE AF)
    Delay Volume :70µL
    Dimensions :15.74D x 10.23W x 6.29 in. H (40 x 26 x 16cm)
    Diode Spacing :1.2nm
    Diodes :512
    Discrete Channels :Two wavelength-selectable channels
    Drift :<1 mAU/hr at 254nm and 546nm using a 50mm LightPipe flow cell
    <0.1 mAU/hr at 254nm and 546nm using a 10mm standard analytical quartz flow cell
    Drift/Unit Temperature :2 x 10-7 RIU/°C
    Drift/Unit Time :2 x 10-7 RIU/h
    Flow Rate Accuracy :±0.5% at set point or +1µL/min whichever is greater
    Flow Rate Resolution :0.1µL/min
    Flowcell Pressure :5kg/cm2 max
    Flowcell Volume :8µL
    Injection Volume :25µL loop standard, up to 1000µL injections with larger loops and syringe
    Integrator Output :0 to 1V (512 x 10-5 RIU/FS)
    Light Source :Pre-aligned Deuterium and Tungsten lamps
    Linearity :<60 x 10-5 RIU
    Maximum Flow Rate :10mL/min
    Maximum Operating Pressure :18,000psi
    Measuring Method :Deflection method
    Minimum Flow Rate :1.0µL/min (isocratic)
    Minimum Sample Volume :1µL can be injected from a 5µL sample
    Needle Height :Programmable in 0.1mm increments. Active vial bottom search selectable on/off
    Pressure Range :0 to 1250 bar (0 to 18,130psi)
    Pressure Resolution :0.01 bar (0.15psi)
    Primary Piston Volume :23µL
    Pulse Dampener :No pulse dampener
    Recommended Flow Rate :0.2 to 3.0mL/min
    Recorder Output :0 to 10mV
    Refractive Index Range :1.00 to 1.75 RIU
    Remote Outputs :Pump Stop, Autosampler Ready, Inject, and Gradient Start
    4 Timed Events
    Reproducibility :<1% RSD (partial loop)
    Residual Pulsation :<0.5 bar amp. (typical)
    Response :FAST/STD/SLOW (3 steps)
    Rise Time :User selectable: 0, 0.1, 0.2, 0.5, 1, 2, 3, 4, or 5 sec
    Sample Carryover :<0.003% chlorhexidine (with two wash solutions) and UV detection (257nm)
    Sample Preparation :Capacity for four 15mL reagent vials for flushing, washing, sample dilution, and/or reagents for sample derivatization
    Scan Rate :User selectable: 4, 5, 6.67, 10 or 20Hz
    Short-Term Noise :<6µAU/cm at 254nm, 2 sec rise time, single wavelength mode using a 50mm LightPipe flow cell
    <20µAU/cm at 254nm, 2 sec rise time, single wavelength mode using a 10mm standard analytical quartz flow cell
    Short-Term Noise (50 mm flow cell) :<6µAU/cm at 254nm at 5Hz (20Hz scan rate)
    <1.5µAU/cm at 254nm at 5Hz (40Hz scan rate); typical
    <0.3µAU/cm at 254nm at 5Hz (80Hz scan rate); typical
    Short-term Noise :2.5 x 10-9 RIU
    Solvent Capacity :Quaternary valve system
    Spectral Bandwidth :6nm
    Syringe Size :100µL
    Optional syringe sizes available (10 to 5000µL)
    Tray Temperature Control :4 to 40°C
    Typical Clean Cycle Time :Less than 1 minute for Fast Cycle
    Variable Volume :Injection Precision: 1.0% RSD at 5µL and higher
    Vial Capacity :324 vials (2.0mL)
    Twelve 96- or 384-well plates
    Six 96- or 384-deep well plates
    Warm-up Time :90 min. required to meet noise and drift specifications
    Wavelength Accuracy :±1nm at 254nm and 656nm
    Wavelength Calibration :Using Holmium oxide solution
    Wavelength Precision :<0.01nm
    Wavelength Range :190 to 800nm
    Deuterium Lamp: 190 to 365nm
    Tungsten Lamp: 366 to 800nm
    Wavelength Resolution :1.2nm (512 pixel array)
    Weight :39.6 lb. (18kg)